Product event summary: the 990063-020 mapping catheter with lot 225623822 was returned and analyzed.Visual inspection was performed, and the tip and loop section were stuck inside the introducer due to tightened torquer.Bond damage was observed at the shaft to the tube fitting.Visual inspection of the loop segment area showed the loop was intact with no apparent issues.No damage was observed along with the tip/loop section of the mapping catheter.Visual inspection of the segment area showed the tubing was kinked/twisted at approximately 3.8 inches from loop distal tip.Visual inspection of the segment area showed the tubing was stuck into the introducer and the tubing was damaged.The insertion compatibility inspection was performed with the test catheter and the returned mapping catheter.No anomalies were identified.As received the mapping catheter was inserted and retracted into the balloon test catheter without any issue.The mapping catheter was intact with no apparent issues.No kinks were observed along with the shaft, or tip/loop section of the mapping catheter.In conclusion, the mapping catheter due to bond damage on the shaft to tube fitting and a kink that was observed on the tip/loop of the tubing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|