• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 1 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 1 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number LOGIC TIBIA PS MOD INSRT SZ 1 11MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 04/28/2023
Type of Reportable Event Serious Injury
Event or Problem Description
During the week of april 17-25, representatives of exactech, inc.And exactech spain conducted an in person site visit with (b)(6) hospital.As a result of that site visit and at exactech's request, the hospital provided a retrospective excel listing of revision related to polyethylene wear that have been performed since 2010.These cases have not previously been reported to exactech as complaints.This patient (549) was implanted with a 02-012-35-1011-inserto tibial logic ps 1, 11 mm, serial number (b)(4) on (b)(6) 2014.The insert was manufactured on 20-mar-2014 and was packaged in a vacuum bag with no evoh.The insert was manufactured on 20-mar-2014 and was six months shelf age at the time of implant.Approximately nine years post implantation, the patient is pending revision for wear of the polyethylene.No additional details are available at this time.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3 the prosthesis wear reported may have been due to third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components.However, this cannot be confirmed as the devices were not available for evaluation and radiographs were not provided.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Event or Problem Description
It was reported that the patient underwent an initial total knee arthroplasty.Subsequently, the patient experienced wear of the tibial insert.As a result, the patient is scheduled for a knee revision on a future date.No further patient impact reported.No further information available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b2, b3, b5, d8, g1, g2, h6 - health effect - clinical code, impact code, investigation findings, investigation conclusions, h7 and h11.B3: unknown date.The prosthesis wear reported may have been due to third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components.However, this cannot be confirmed as the devices were not available for evaluation and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 1 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
ashley harris
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key16950602
Report Number1038671-2023-01007
Device Sequence Number6850837
Product Code JWH
UDI-Device Identifier10885862001696
UDI-Public10885862001696
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2022
Device Model NumberLOGIC TIBIA PS MOD INSRT SZ 1 11MM
Device Catalogue Number02-012-35-1011
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2023
Supplement Date Received by Manufacturer06/30/2023
02/27/2026
Initial Report FDA Received Date05/17/2023
Supplement Report FDA Received Date07/28/2023
03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/20/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0021-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-