• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752200
Device Problem Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2023
Event Type  malfunction  
Event Description
A physician reported that the aspiration was poor in ultrasound and irrigation or aspiration mode during the cataract surgery.The surgery was completed without product replacement.There was no patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.No physical sample has been returned for evaluation; therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16951088
MDR Text Key315548065
Report Number1644019-2023-00579
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657522002
UDI-Public00380657522002
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752200
Device Lot Number150HC5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-