Model Number 55740004530
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Paraplegia (2448)
Date of Event 04/25/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having t4/l2 correctio n/fixation, insertion of ps under mazor for ais.It was reported that after insertion of the ps, it was confirmed that the t9 left, t5 right, and t6 right of the ps were deviated from the spinal canal, and then they were removed.The patient experienced paralysis in the right lower limb.There was a delay in the procedure by less than 60 mins.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
B2.The patient experienced paralysis in the right lower limb.H3.Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key16951384
Report Number1030489-2023-00373
Device Sequence Number12515751
Product Code KWP
UDI-Device Identifier00613994996718
UDI-Public00613994996718
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202328
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55740004530
Device Catalogue Number55740004530
Device Lot NumberH5816189
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2023
Initial Report FDA Received Date05/18/2023
Date Device Manufactured11/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability;
Patient SexUnknown
-
-