INMODE LTD BODYTITE Back to Search Results
Model Number AG604881A
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Full thickness (Third Degree) Burn (2696)
Date of Event 04/21/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Despite more requests to date we have not obtained the information relevant to the treatment performed by the doctor.However, excluding problems of a technical nature, it can be deduced from the analysis of the photos that an incorrect handling of the cannula was performed with ascent to the surface in the distal tip and excessive energy delivery contrary to what is instructed in the user manual.
 
Event or Problem Description
2 full-thickness burns (3rd degree) of oval shape about 1 cm in diameter each.Anatomical location not clear in the photo, presumably on the right side of the abdomen.One of the lesions appears devoid of dermal tissue while the other has a necrotic black crust.The wounds require prophylactic medical care in order to prevent secondary infection and to boost the healing process.At the moment, scar formations in correspondence with the lesions cannot be excluded.Still awaiting further details from the clinic.
 
Additional Manufacturer Narrative
Follow up report 07/02/2023.1.User error - the root cause leads to an error in the technique of superficial approach of the cannula to the access port, emitting an excessive energy tantum in the superficial dermal layer.Investigation conclusions: maneuvering error - excess surface energy in the vicinity of the access port.2.Retrain summary isr added.3.Final test added.
 
Event or Problem Description
2 full-thickness burns (3rd degree) of oval shape about 1 cm in diameter each.Anatomical location not clear in the photo, presumably on the right side of the abdomen.One of the lesions appears devoid of dermal tissue while the other has a necrotic black crust.The wounds require prophylactic medical care in order to prevent secondary infection and to boost the healing process.At the moment, scar formations in correspondence with the lesions cannot be excluded.Still awaiting further details from the clinic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BODYTITE
Common Device Name
BODYTITE
Manufacturer (Section D)
INMODE LTD
tavor building (pob 533)
shaar yokneam
yokneam northern, north 20692 06
IS  2069206
Manufacturer (Section G)
INMODE LTD.
tavor building (pob 533)
shaar yokneam
yokneam northern, north 20692 06
IS   2069206
Manufacturer Contact
eran krieger
tavor building (pob 533)
shaar yokneam
yokneam northern, north 20692-06
IS   2069206
MDR Report Key16952371
Report Number3010511300-2023-00226
Device Sequence Number7012866
Product Code GEI
UDI-Device Identifier07290016633405
UDI-Public07290016633405
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K182325
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Remedial Action Inspection,Patient Monitoring,Other,Notification
Type of Report Initial,Followup
Report Date (Section B) 07/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAG604881A
Device Catalogue NumberAG604881A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2023
Supplement Date Received by Manufacturer06/06/2023
Initial Report FDA Received Date05/18/2023
Supplement Report FDA Received Date07/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age40 YR
Patient SexPrefer Not To Disclose
-
-