Model Number HX-110QR
Medical Device Problem Code Device Reprocessing Problem (1091)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was used by the facility for two months after the delivery of the device.So the number of cases, for which the device was used is unknown.The device was returned.And customer allegation was confirmed.A black foreign substance was found on the gauze.The appearance of the hook was checked and no bends or deformities was found.There were no problems with hook extrusion and retraction.It was found, that the coating (black part) of the operating wire was peeling off.The investigation is ongoing.A supplemental will be submitted on completion of investigation or if any additional information is received.
 
Event or Problem Description
The customer reported to olympus, when the user moved the operating handle of the rotatable clip fixing device.Black sand like foreign material came out of the coil sheath from the tip of the device.The issue was found during reprocessing.The device was washed again, but the same issue occurred again.After the foreign material was found, the device was not used.And there were no reports of any health damage or patient harm associated with the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was manufactured in september 2022.However, the specific date is unknown.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the foreign material and peeled coating could not be determined.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿before use, check that the hook and coil sheath are not corroded (microscopic holes, dents, discoloration, etc.) and that there is no abnormality in the operational feel.Using this product with corroded hook or coil sheath may cause the hook or coil sheath to fall off.During use, always check the endoscopic image to confirm that there is no abnormality in the tip and that there is no abnormality in the operability.In the unlikely event that the hook or coil sheath falls off during use, please retrieve it with grasping forceps.Do not squeeze, wipe, or rub the rotating clip device.It may lead to damage or deterioration of the function of the rotating clip device.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTATABLE CLIP FIXING DEVICE
Common Device Name
ROTATABLE CLIP FIXING DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16952890
Report Number9614641-2023-00690
Device Sequence Number7012881
Product Code PKL
UDI-Device Identifier04953170200205
UDI-Public04953170200205
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K013066
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHX-110QR
Device Lot Number29K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2023
Supplement Date Received by Manufacturer06/13/2023
Initial Report FDA Received Date05/18/2023
Supplement Report FDA Received Date07/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NON-OLYMPUS ULTRASONIC CLEANER.
Patient SexUnknown
-
-