COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM STANDARD; STENT, SUPERFICIAL FEMORAL ARTERY
|
Back to Search Results |
|
| Model Number |
EVD35-08-040-120 |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A physician was attempting to use an everflex entrust for the treatment of a calcified lesion with 70% stenosis in the proximal region of the superficial femoral artery.The artery was mildly tortuous and moderately calcified.The device was prepped as per the ifu with no issues identified.The lesion was pre-dilated.The device did not pass through a previously deployed stent.The thumbscrew was checked for securement prior to the procedure and it was removed once the stent was positioned at the lesion and ready for deployment.It was reported the surgeon approached from the patient¿s brachial artery, first used a 6mm balloon to pre-dilate the lesion, and then used a 7mm balloon to dilate, and then attempted to use an 8mmx 40 stent to complete the treatment of the lesion, but it was found that after the stent delivery system was sent to the lesion and when ready to deploy, the rotating thumbwheel got stuck after rolling a few times and could no longer be operated, so the whole body was withdrawn from the patient's body.The device was safely removed from the patient.A little tip of the stent was exposed upon removal.The stent was replaced in order to complete the treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis the device was returned with the red safety lock pin removed, (photo 2).The stent was not returned with the device.The device was returned with a kink on the outer sheath at approx.13.2cm from the strain relief at end of outer blue sheath, an in house 0.035¿ guidewire was advanced through the device, but it could not pass the kink the handle was dismantled, and the pull cable was found to be intact.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|