• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 40MM VIT E LINER +0MM; SHOULDER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 40MM VIT E LINER +0MM; SHOULDER, PROSTHESIS Back to Search Results
Model Number 00-4350-040-00
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 04/17/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was revised due to baseplate disassociation, liner wear and pain. no additional patient consequences were reported.  attempts have been made and additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565 -2023 -01349.D10: base plate uncemented 25 mm post length cat: 00436202500 lot: 64610507.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Upon further review, it was determined that the disassociation is between the glenosphere and the baseplate.The liner is not related to the event.This event will be investigated on medwatch#s:0001822565 -2023 -02618; 0001822565 -2023 -01349.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
40MM VIT E LINER +0MM
Common Device Name
SHOULDER, PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16953156
Report Number0001822565-2023-01360
Device Sequence Number20586293
Product Code HSD
UDI-Device Identifier00889024269422
UDI-Public(01)00889024269422(17)270501(10)65226145
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K133378
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00-4350-040-00
Device Catalogue Number00435004000
Device Lot Number65226145
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2023
Supplement Date Received by Manufacturer09/20/2023
Initial Report FDA Received Date05/18/2023
Supplement Report FDA Received Date09/22/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight99 KG
-
-