• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS CORPORATION - MONTGOMERY LB53 LIGHT HANDLE COVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CORPORATION - MONTGOMERY LB53 LIGHT HANDLE COVER Back to Search Results
Model Number STERILE LIGHTHANDLE COVER
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Steris has confirmed the reported issue.The polyblend plastic material that is used to form the light handle cover was sourced from a secondary supplier beginning in (b)(6) 2022 and may not meet performance specifications.Steris initiated a recall (removal) of the affected product on (b)(6) 2023.
 
Event or Problem Description
The user facility reported via medwatch report (2601800000-2023-8003) that during a patient procedure their light handle cover detached and fell.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LB53 LIGHT HANDLE COVER
Common Device Name
LIGHT HANDLE COVER
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
MDR Report Key16954208
Report Number1043572-2023-00103
Device Sequence Number16969295
Product Code FTA
UDI-Device Identifier00724995150136
UDI-Public00724995150136
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSTERILE LIGHTHANDLE COVER
Device Catalogue NumberLB53
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2023
Supplement Date Received by Manufacturer04/18/2023
Initial Report FDA Received Date05/18/2023
Supplement Report FDA Received Date06/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-