| Model Number |
1013465-080 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a moderately calcified, 75% stenosed lesion in the tibial artery.A 3.0mmx80mm/150cm armada 18 balloon dilatation catheter (bdc) was advanced to the lesion without issue and inflated one time to 8 atmospheres (atm) when the balloon ruptured.The bdc was removed, and a 2.5 abbott balloon was used to successfully complete the procedure.There were no reported adverse patient effects and no clinically significant delays in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaint appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|