| Model Number |
6144ICLL |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported they experienced a foley catheter dislodgement that appeared to be related to a leaking balloon.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Samples were not received for the investigation.A device history record review could not be performed because a lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the reported issue and determine the root cause, therefore a corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.If a sample is received at a later date, this complaint will be reopened for further investigation.This complaint will be used for tracking and trending purposes.
|
| |
|
Search Alerts/Recalls
|