EXACTECH, INC. CC TIBIAL INSERT SZ 3, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
208-23-11 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
04/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending investigation.H7: z-0019-2022.
|
| |
|
Event or Problem Description
|
|
During the week of (b)(6), representatives of exactech, inc.And exactech spain conducted an inperson site visit with (b)(6) hospital.As a result of that site visit and at exactech's request, the hospital provided a retrospective excel listing of revision related to polyethylene wear that have been performed since 2010.These cases have not previously been reported to exactech as complaints.This patient (b)(6) was implanted with a 208-23-11-inserto tibial cc 3,11mm, serial number (b)(6) on (b)(6) 2013.The insert was manufactured on 7/27/2009 and was packaged in a vacuum bag with no evoh.The insert was manufactured on 7/27/2009 and was 3.74 years shelf age at the time of implant.Pending revision for wear of the polyethylene.No additional details are available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the prosthesis wear reported may have been due to third body wear, patient-related conditions, instability, malalignment between the femoral and tibial components, and/or being implanted for over 10 years.However, this cannot be confirmed as the devices were not available for evaluation and radiographs were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The prosthesis wear reported may have been due to third body wear, patient-related conditions, instability, malalignment between the femoral and tibial components, and/or being implanted for over 10 years.However, this cannot be confirmed as the devices were not available for evaluation and radiographs were not provided if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|