MEDTRONIC NEUROMODULATION INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97725
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: product id 977d260 lot# serial# (b)(6) implanted: explanted: product type screening device product id 977d260 lot# serial# (b)(6) implanted: explanted: product type screening device section d information references the main component of the system.Other relevant device(s) are: product id: 977d260, serial/lot #: (b)(6), ubd: 24-mar-2027, udi#: (b)(4) product id: 977d260, serial/lot #: (b)(6), ubd: 24-mar-2027, udi#: (b)(4) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a manufacturer representative regarding a trial patient who was using an external neurostimulator (ens).It was reported that when mapping the patient in the procedure room, lead connection was checked and showed "disconnect" on several electrodes even after removing/replacing them from the ens.Impedances were listed as 40,000 for multiple electrodes.The leads were tested with another ens and the problem continued.They removed both leads and implanted new leads.The issue was resolved.
 
Additional Manufacturer Narrative
H3: product analysis #291129578:analysis information -- 2023-07-12 13:50:15 cst pli# 10 product id# 97725 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and functional testing.The returned device passed all testing in the laboratory and no anomalies were identified.Analysis information -- 07/25/2023 10:28:43 cst pli# 20 product id# 977d260 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned device passed all testing in the laboratory and no anomalies were identified.Analysis information -- 07/25/2023 09:14:41 cst pli# 30 product id# 977d260 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned device passed all testing in the laboratory and no anomalies were identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16963217
Report Number2182207-2023-00907
Device Sequence Number14761901
Product Code LGW
UDI-Device Identifier00763000278762
UDI-Public00763000278762
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2024
Device Model Number97725
Device Catalogue Number97725
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2023
Supplement Date Received by Manufacturer07/25/2023
Initial Report FDA Received Date05/19/2023
Supplement Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Patient Age66 YR
Patient SexUnknown
-
-