• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AECJ-502
Medical Device Problem Codes Fracture (1260); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2023 a 19mm sjm masters series valve w/ expanded cuff was chosen for implant.The patient's aortic annulus was sized 19mm via a mechanical sizer set.After the device was implanted, the operator attempted to rotate the valve in situ, felt resistance, and was unable to rotate when one of the leaflets had fractured into three pieces.It was reported no other instruments encountered the valve at the time of fracture and all of the leaflet fractured pieces were removed from the patient.It was also reported only the holder handle was used to rotate the valve, the holder hander was fully inserted into the orifice at a 90 degree angle, and the valve was in a closed position when rotated.A decision was made to explant the valve and a new 19mm sjm masters series valve expanded cuff was implanted in the patient with no adverse patient consequences.There was no clinically significant delay in the procedure due to this event.The patient remained hemodynamically stable throughout the procedure and their status was reported as stable.
 
Additional Manufacturer Narrative
The reported event of a dislodged leaflet was confirmed.One leaflet was fractured and dislodged from the orifice.No other damage was noted to the returned valve.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.While the root cause of the leaflet dislodgement could not be conclusively determined, there was no evidence of material defect in the carbon coating that may have caused or contributed to the dislodged leaflet or orifice damage.The damage may have been caused by some external force applied to the valve, which overstressed the carbon material.Please note, per the masters valve instructions for use, "valve rotation: using the valve holder/rotator and an sjm valve holder handle, rotate the valve in situ to the desired position.The valve should rotate freely.If resistance is noted, the valve holder/ rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES VALVE EXPANDED CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16969670
Report Number2135147-2023-02229
Device Sequence Number12125666
Product Code LWQ
UDI-Device Identifier05414734053877
UDI-Public05414734053877
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/07/2026
Device Model Number19AECJ-502
Device Catalogue Number19AECJ-502
Device Lot Number8128805
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2023
Supplement Date Received by Manufacturer07/06/2023
Initial Report FDA Received Date05/19/2023
Supplement Report FDA Received Date07/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
Patient Weight57 KG
-
-