|
EXACTECH, INC. CR TIBIAL INSERT SZ 1D, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
CR TIBIAL INSERT SZ 1D, 9MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
04/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
During the week of april 17-25, representatives of exactech, inc.And exactech spain conducted an in person site visit with (b)(6) hospital.As a result of that site visit and at exactech's request, the hospital provided a retrospective excel listing of revision related to polyethylene wear that have been performed since 2010.These cases have not previously been reported to exactech as complaints.This patient (b)(6) was implanted with a 200-51-09-inserto tib cr 1d, 9mm, serial number: (b)(6) on (b)(6) 2013.The insert was manufactured on (b)(6) 2011 and was packaged in a vacuum bag with no evoh.The insert was manufactured on 16-dec-2011 and was 1.4 years shelf age at the time of implant.Approximately ten years post implantation, the patient is pending revision for wear of the polyethylene.No additional details are available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3 the prosthesis wear reported may have been due to third body wear, patient-related conditions, instability, malalignment between the femoral and tibial components, and/or being implanted for over 10 years.However, this cannot be confirmed as the devices were not available for evaluation and radiographs were not provided.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b2, b3, d8, h6 - health effect - clinical code, health effect - impact code, investigation findings, investigation conclusion and h11.The revision reported may have been due to third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components, which led to prosthesis wear.However, this was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|