| Model Number |
372010-02 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the endoscope associated with this complaint and completed its investigation.Failure analysis (fa) confirmed the reported issue as the endoscope was received with missing distal window resulting in fluid invasion and subsequent major damage to optical components.The scope window was completely missing and fragments of adhesive of the distal window were also missing.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the lens on distal tip of the endoscope was missing.The site reported that the "mirrors might not be in proper orientation" and that there appeared to be some damage on right the side.The procedure was completed with no reported injury.
|
| |
|