HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO PROCESSOR (FOR VIVIDEO) Back to Search Results
Model Number CP-1000
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/08/2023
Event Type  malfunction  
Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the image missing.Based on the result, we concluded that it was caused due to the excessive shock applied on the imaging device.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is not during procedure.There was no report of patient harm.Video image failure(blackout).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO PROCESSOR (FOR VIVIDEO)
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key16971779
MDR Text Key315664876
Report Number9610877-2023-54546
Device Sequence Number1
Product Code EOB
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K171011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCP-1000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/08/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-