• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH MRH KNEE FEM XS RHT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH MRH KNEE FEM XS RHT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 6481-1-101
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Patient reported for revision surgery due to query of failed mrh component.Confirmed during surgery that mrh distal femur and stem extender had broken and detached.Impact: revision surgery convert mrh to distal femur gmrs.Procedure: mrh to proximal femur gmrs reconstruction revision.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a mrh femoral component was reported.The event was confirmed via visual inspection of the provided photographs of the device.Method & results: product evaluation and results: the reported device was not returned however photographs were provided for review.Visual inspection of the provided photographs of the devices indicate that the devices have been recently explanted and are covered in blood/tissue and bone cement.A portion of the threaded junction of the femoral component appears to have fractured off.It also appears that a fractured portion of the stem extender remains threaded into the femoral component.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to fracture at the stem/femoral component interface.The reported device was not returned however photographs were provided for review.Visual inspection of the provided photographs of the devices indicate that the devices have been recently explanted and are covered in blood/tissue and bone cement.A portion of the threaded junction of the femoral component appears to have fractured off.It also appears that a fractured portion of the stem extender remains threaded into the femoral component.Further information such as return of the devices, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient reported for revision surgery due to query of failed mrh component.Confirmed during surgery that mrh distal femur and stem extender had broken and detached.Impact: revision surgery convert mrh to distal femur gmrs procedure: mrh to proximal femur gmrs reconstruction revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MRH KNEE FEM XS RHT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key16973536
Report Number0002249697-2023-00568
Device Sequence Number20495356
Product Code KRO
UDI-Device Identifier07613327045123
UDI-Public07613327045123
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K994207
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number6481-1-101
Device Catalogue Number64811101
Device Lot NumberCY94L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/26/2023
Supplement Date Received by Manufacturer08/03/2023
Initial Report FDA Received Date05/22/2023
Supplement Report FDA Received Date08/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/31/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age19 YR
Patient SexMale
-
-