LDR MÉDICAL ANCHORING CLIP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number MC001T
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994)
Type of Reportable Event Serious Injury
Event or Problem Description
A patient reported inflammation and chronic pain due to the metals or plastic components in two mc+ cages and plates.The constructs remain implanted.This is report one of four for this event.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3004788213-2023-00048 through 3004788213-2023-00051.
 
Additional Manufacturer Narrative
H6: additional method code: 4109.Corrections in g1.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the device was not returned and no photos were provided, so an evaluation is unable to be performed.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
A patient reported inflammation and chronic pain due to the metals or plastic components in two mc+ cages and plates.The constructs remain implanted.This is report one of four for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANCHORING CLIP
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
sabrina abla
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key16973551
Report Number3004788213-2023-00048
Device Sequence Number17823495
Product Code NKB
UDI-Device Identifier03662663022361
UDI-Public(01)03662663022361(17)210101(10)526724/A
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K121103
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/01/2021
Device Catalogue NumberMC001T
Device Lot Number526724/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2023
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/22/2023
Supplement Report FDA Received Date06/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/03/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-