• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-450-14
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that two pipeline's failed to open in the distal section.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm in the left internal carotid artery (ica) with a max diameter of 9 mm and a 4 mm neck diameter.The landing zone was 3.8 mm distally and 4.3 mm proximally.It was noted the patient's vessel tortuosity was moderate.Dual antiplatelet treatment was administered.The pru level was 125.The angiographic result post procedure showed successful flow diversion.It was reported that the physician discarded 2 pipelines because wasn't comfortable deploying in that location. pipelines fish mouthed slightly in patient but opened as normal on back table.Physician replaced with 2 pipelines from my trunk stock that deployed successfully.It was reported both of the pipelines failed to open in the distal section.The pipeline was positioned in a distal bend.More than 50% of the pipeline had been deployed when it failed to open.The pipeline was resheathed more than two times.There were no additional steps or other devices required to open the pipeline.The pipeline was removed from the patient by resheathing and removing with the microcatheter.The pipeline was not used for an indication that is off-label.The reported pipeline and any accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or further complications were reported as a result of this event.Ancillary devices include a bmx 81 sheath.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key16973908
Report Number2029214-2023-00824
Device Sequence Number2317811
Product Code OUT
UDI-Device Identifier00763000284510
UDI-Public00763000284510
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/10/2025
Device Model NumberPED2-450-14
Device Catalogue NumberPED2-450-14
Device Lot NumberB337206
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2023
Initial Report FDA Received Date05/22/2023
Date Device Manufactured02/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age82 YR
Patient SexFemale
-
-