|
Zimmer biomet complaint number (b)(4).Zimvie received one (1) ilrght, (certain titanium large hexed screw) for evaluation.Visual evaluation was performed, a fracture has been identified at the threads.Dhr review was completed for the subject lot number 1255182.No deviations or non-conformance's, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1255182 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : screw based on the investigation and risk management file review, the most likely root cause determined from the investigation were material selection of the product is not adequate to withstand occlusal forces.Therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the threads.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|