This report is being submitted to relay additional information and device evaluation.Zimvie received the reported implant for evaluation.Visual evaluation was performed, a fracture has been identified on the implants body.Dhr review was completed for the subject lot number 1241937.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1241937 for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation were missing or confusing instructions for use and clinician selects implant that is unable to resist long-term occlusal loading.Therefore, based on the available information, device malfunction did occur.The fracture has been identified on the body.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|