| Model Number |
9-AVP2-016 |
| Medical Device Problem Codes |
Fracture (1260); Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2023, a 16mm amplatzer vascular plug ii was chosen for implant with an unknown delivery system.During procedure, it was noted via intraoperative angiography there was a hole in the device.A decision was made to remove the device and replaced with another amplatzer vascular plug ii.The first device was never released from the delivery cable while inside the patient.There was no fracture/breakage of the nitinol wires on the device and there was no note of packaging damage prior to use.It was reported the patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in the procedure due to this event and the patient status was reported as stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of a hole being noted in the device via intraoperative angiography.One image was received appearing the show the hole in the device.The device was returned to abbott for analysis.The reported event of a hole in device was confirmed; the forming hole is a result of the manufacturing process.The investigation confirmed the device met specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event is consistent with a forming hole which is considered to be a normal and acceptable characteristic of the device.There were no complaints associated with any other devices from the lot.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|