| Model Number |
CCW0T3 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implant procedure, it was noted that, there was damage of iol.The lens left in the patients eye.Additional information ahs been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The complainant indicates the use of company cartridge with company lens model, this is not a qualified company product combination.The reported complaint was not observed as no sample was returned for analysis; however, based on the information provided by the customer, the most likely root cause is failure to follow instructions for use (ifu), as the surgeon states the use of a non-qualified combination.A company cartridge is not qualified for use with company lens of 28.0 diopter.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Based on the results from the product, batch and sterilization history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|