This report is being supplemented to provide additional information based on the legal manufacturer's final investigation, the device evaluation, and corrections to d9 and h3.The device evaluation and the information in d9 and h3 were inadvertently omitted from the initial medwatch report.Please see updates to d9, h3, h6 and h10.The device was returned to olympus for inspection, and the customer's complaint was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, it¿s likely the sheathing tube broke due to excessive force such as: bending, jamming, kinking, and/or excessive frequent use.The final root cause of this event was unable to be identified.Olympus will continue to monitor field performance for this device.
|