Model Number Dual-Lumen Silicone
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Utmd is reporting this event because the hospital staff stated that a interventional radiology was performed to remove the remaining catheter from the patient.The mgr nicu peds at user facility stated that the patient was stable and there was no known long term effects to the patient.After investigation at utmd, the returned catheter had been damaged with sharp instrument and then torn.The returned catheter was pull tested at utmd and met specifications.The ifu has cautions indicating to take care not to damage the soft silicone catheter tubing during use.(see ifu 58078 rev 022322).
 
Event or Problem Description
On (b)(6) 2023.The catheter broke during removal.The remaining piece remained internal and the infant subsequently had to be shipped to the higher level of care for removal.The remaining catheter was removed at another facility with interventional radiology.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UMBILI-CATH
Common Device Name
UMBILICAL CATHETER
Manufacturer (Section D)
UTAH MEDICAL PRODUCTS, INC
7043 south 300 west
midvale UT 84047
Manufacturer (Section G)
UTAH MEDICAL PRODUCTS,INC
7043 south 300 west
midvale UT 84047
Manufacturer Contact
ben shirley
7043 south 300 west
midvale, UT 84047
8015661200
MDR Report Key16987272
Report Number1718873-2023-00001
Device Sequence Number6791002
Product Code FOS
UDI-Device IdentifierH67142735050
UDI-PublicH67142735050
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K940953
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberDual-Lumen Silicone
Device Catalogue Number4273505
Device Lot Number1223017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age7 DA
Patient SexUnknown
Patient Weight2 KG
-
-