• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER LEIBINGER FREIBURG L-PLATE, 100 DEGREE, 8MM BAR, LEFT, GSP, 6 HOLE, MIDFACE; IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER LEIBINGER FREIBURG L-PLATE, 100 DEGREE, 8MM BAR, LEFT, GSP, 6 HOLE, MIDFACE; IMPLANT Back to Search Results
Model Number 55-07764
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 07/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the company representative that upon re-operation to explant the plates due to the maxilla shifting, it was discovered that the plates were fractured.
 
Additional Manufacturer Narrative
The device has not yet been received at the manufacturer for testing.An evaluation will be conducted upon receipt of the device, and a follow-up report will be submitted after the quality investigation is complete.
 
Additional Manufacturer Narrative
The reported event (plate breakage) could not be confirmed as the plates were not returned for investigation.It has been reported that 3 plates were broken.As a result, a reoperation was performed.The sales rep mentioned that the patient had initial surgery in (b)(6) 2022.He was compliant to post op care instructions and attended post op follow-ups regularly.The surgeon noticed movement of the maxilla in (b)(6) 2023.He also observed non-union prior to the reported plate breakages.The post op images were reviewed.The design team did not observe any broken plates.Stryker¿s medical expert stated that non-union will cause plate breakage.Union usually occurs 4-5 months after surgery.Based on statistical evaluation, there is no indication for an incorrectly working product or any design, material or manufacturing related issue.
 
Event or Problem Description
It was reported by the company representative that upon re-operation to explant the plates due to the maxilla shifting, it was discovered that the plates were fractured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-PLATE, 100 DEGREE, 8MM BAR, LEFT, GSP, 6 HOLE, MIDFACE
Common Device Name
IMPLANT
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
rylee koole
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
7614888208
MDR Report Key16987430
Report Number0008010177-2023-00027
Device Sequence Number12475893
Product Code JEY
UDI-Device Identifier04546540366139
UDI-Public04546540366139
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K022185
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model Number55-07764
Device Catalogue Number55-07764
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2023
Supplement Date Received by Manufacturer05/04/2023
Initial Report FDA Received Date05/23/2023
Supplement Report FDA Received Date09/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-