| Model Number |
408310 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Innovative health, llc became aware on (b)(6) 2023 of a report from (b)(6) hospital - (b)(6) on an agilis nxt steerable introducer.The device was reported to have a one way valve and hub found to be faulty during a procedure.Innovative health received the device for evaluation on 5/4/2023.Upon investigation, the hemostasis cap was identified to be detached from the sideport hub.No injury was reported.
|
| |
|
Event or Problem Description
|
|
This device was reported to have a one-way valve and hub found to be faulty during a procedure.No injuries were reported.
|
| |
|