Model Number 408310
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Innovative health, llc became aware on (b)(6) 2023 of a report from (b)(6) hospital - (b)(6) on an agilis nxt steerable introducer.The device was reported to have a one way valve and hub found to be faulty during a procedure.Innovative health received the device for evaluation on 5/4/2023.Upon investigation, the hemostasis cap was identified to be detached from the sideport hub.No injury was reported.
 
Event or Problem Description
This device was reported to have a one-way valve and hub found to be faulty during a procedure.No injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Common Device Name
STEERABLE INTRODUCER
Manufacturer (Section D)
INNOVATIVE HEALTH, LLC.
1435 north hayden road
suite 100
scottsdale AZ 85257
Manufacturer (Section G)
INNOVATIVE HEALTH, LLC.
1435 north hayden road
suite 100
scottsdale AZ 85257
Manufacturer Contact
rebecca sturm
1435 north hayden road
suite 100
scottsdale, AZ 85257
MDR Report Key16988675
Report Number3011610434-2023-00007
Device Sequence Number6693741
Product Code PNE
UDI-Device Identifier10841898120275
UDI-Public10841898120275
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K170311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2024
Device Model Number408310
Device Catalogue Number408310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2023
Initial Report FDA Received Date05/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-