Catalog Number AA01
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have received limited information from the end user.We are continuing to follow up.As of 1/01/2021 there have been 3 similar incidences with (b)(4) cauteries sold.Once we have received additional information, a follow up report will be submitted.
 
Event or Problem Description
The customer alleged that the cauteries are getting hot and sparking when in use.
 
Additional Manufacturer Narrative
Received the complaint through an mdpr in canada.Product was purchased on order number 958277 on (b)(6) 2022 by ndc.Product was shipped on invoice number 1197108 on 11/15/2022 line 10.Product was sold to canada distributor on po# 4600078342.Product was sold to end user on an unknown date.Customer stated they were attempting to use the devices, but for 3 eaches, the device was causing sparks when held initially active.Devices were received through the rma room on 6/1/2023.Devices were in good overall condition, with the tips slightly bent.One device had the button pushed in and broken.2 devices were received with corrosion in the battery pack, preventing further activation.1 devices returned would not activate at all, as the button had fallen in.This is not representative of the complaint from the customer, and we are unaware if the items were in this state (corrosion) prior to international return shipment.Customer was asked again about information for use of the device on (b)(6) 2023 since testing of the devices could not confirm the complaint.Customer responded stated they have been a long time user of the device with no changes in environment or users.Customer stated they had this issue out of box, where the three devices were selected for use, but sparked when activated.Initial complaint case was unable to be confirmed, as the products returned were not in operating condition.Other issues (battery corrosion) was confirmed from the returned units.True root cause for the issues experienced by the customer were unable be confirmed with accuracy.It may be possible this was an issue caused by a failed/damaged battery.Based on the above information, no further actions are required and this can be seen as the final report.If additional information is obtained that alleges any additional patient involvement or additional information pertinent to the investigation, a subsequent follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOVIE
Common Device Name
CAUTERY
Manufacturer (Section D)
BOVIE MEDICAL CORP
5115 ulmerton road
clearwater FL 33760
Manufacturer (Section G)
BOVIE MEDICAL CORP
5115 ulmerton road
clearwater FL 33760
Manufacturer Contact
brandi meath
3034 owen drive
antioch, TN 37013
6159645290
MDR Report Key16988860
Report Number3007208013-2023-00021
Device Sequence Number6872113
Product Code HQP
UDI-Device Identifier10607151011014
UDI-Public10607151011014
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K121441
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2024
Device Catalogue NumberAA01
Device Lot Number0522X
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2023
Supplement Date Received by Manufacturer05/01/2023
Initial Report FDA Received Date05/23/2023
Supplement Report FDA Received Date06/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-