• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II ISRT XLPE DD LT SZ 5-6 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II ISRT XLPE DD LT SZ 5-6 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74025761
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint reference number: (b)(4).
 
Event or Problem Description
It was reported that during a tka surgery, when opening the box, the jrny ii isrt xlpe dd lt sz 5-6 9mm fell to the floor and it was found that the double sterile packages were open about 5cm.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
The associated packaging was returned and evaluated.The visual inspection revealed the sterile packaging is open on one side.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According to the packaging sequence, the component should be packed in an inner pouch and sealed, then it should be placed into outer pouch and sealed for sterilization, lastly it is inserted into carton, sealed and shrink wrap.A review made by the quality engineering team revealed that this device is subjected to multiple visual and dimensional inspection during processing, and this type of damage has a high likelihood of detection during these inspections.Additionally, this is the only complaint for this batch, which was manufactured in 2017 and has a high likelihood of having additional parts from the batch being used in surgery.Based on the assessment, the defect occurred in-transit and is not related to a manufacturing defect.Seal for both the inner and outer pouch were ruptured in the same location.The location is on the left-side of the pouches when looking down on the part.This location is not where heat seal takes place in production and does not indicate a manufacturing discrepancy.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II ISRT XLPE DD LT SZ 5-6 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16989126
Report Number1020279-2023-01111
Device Sequence Number6872124
Product Code JWH
UDI-Device Identifier00885556098417
UDI-Public00885556098417
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K113482
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74025761
Device Catalogue Number74025761
Device Lot Number17HM13230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2023
Supplement Date Received by Manufacturer06/23/2023
Initial Report FDA Received Date05/23/2023
Supplement Report FDA Received Date06/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-