SMITH & NEPHEW, INC. JRNY II ISRT XLPE DD LT SZ 5-6 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
74025761 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a tka surgery, when opening the box, the jrny ii isrt xlpe dd lt sz 5-6 9mm fell to the floor and it was found that the double sterile packages were open about 5cm.The procedure was resumed, after a non-significant delay, with a s+n back-up device.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated packaging was returned and evaluated.The visual inspection revealed the sterile packaging is open on one side.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According to the packaging sequence, the component should be packed in an inner pouch and sealed, then it should be placed into outer pouch and sealed for sterilization, lastly it is inserted into carton, sealed and shrink wrap.A review made by the quality engineering team revealed that this device is subjected to multiple visual and dimensional inspection during processing, and this type of damage has a high likelihood of detection during these inspections.Additionally, this is the only complaint for this batch, which was manufactured in 2017 and has a high likelihood of having additional parts from the batch being used in surgery.Based on the assessment, the defect occurred in-transit and is not related to a manufacturing defect.Seal for both the inner and outer pouch were ruptured in the same location.The location is on the left-side of the pouches when looking down on the part.This location is not where heat seal takes place in production and does not indicate a manufacturing discrepancy.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|