• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. CR TIBIAL INSERT SZ4, 9MM; SEE H10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CR TIBIAL INSERT SZ4, 9MM; SEE H10 Back to Search Results
Model Number CR TIBIAL INSERT SZ4, 9MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/01/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
During the week of april 17-25, representatives of exactech, inc.And exactech spain conducted an inperson site visit with ubarmin hospital.As a result of that site visit and at exactech's request, the hospital provided a retrospective excel listing of revision related to polyethylene wear that have been performed since 2010.These cases have not previously been reported to exactech as complaints.This patient ((b)(6)) was implanted with a 200-24-09-inserto tibial cr 4, 9mm, serial number (b)(6) on (b)(6) 2014.The insert was manufactured on 3/26/2013 and was packaged in a vacuum bag with no evoh.The insert was manufactured on 3/26/2013 and was 1 year shelf age at the time of implant.On (b)(6) 2014, approximately 1.0 years post implantation, the patient underwent revision for wear of the polyethylene.No additional details are available at this time.
 
Additional Manufacturer Narrative
H3: the revision reported may have been due to third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components, which led to prosthesis wear.However, this cannot be confirmed as the device was not available for evaluation and images of the explanted device/radiographs were not provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d1, h6 - investigation findings, investigation conclusion, h7 and h11 the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A definitive root cause was unable to be determined; however, may have resulted from third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components, which led to prosthesis wear.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CR TIBIAL INSERT SZ4, 9MM
Common Device Name
SEE H10
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key16989323
Report Number1038671-2023-01099
Device Sequence Number6828859
Product Code JWH
UDI-Device Identifier10885862040299
UDI-Public10885862040299
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K932776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2021
Device Model NumberCR TIBIAL INSERT SZ4, 9MM
Device Catalogue Number200-24-09
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2023
Supplement Date Received by Manufacturer07/18/2023
03/29/2026
Initial Report FDA Received Date05/23/2023
Supplement Report FDA Received Date08/08/2023
04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/26/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-