• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX42MM STERILE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER + ADDITIVE, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX42MM STERILE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER + ADDITIVE, CEMENTED Back to Search Results
Model Number DWJ742
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/27/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.H3 other text : device disposition is unknown.
 
Event or Problem Description
There was a revision surgery caused by an infection.Additionally, all ten implants were removed.The nurse still had the central screw within the baseplate and the reversed insert was unattached from the reversed tray.
 
Additional Manufacturer Narrative
The reported event was not confirmed since the device was not returned for evaluation and no other evidences were provided.A device inspection was not possible since the affected device was not returned and no other evidences were provided for investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
There was a revision surgery caused by an infection.Additionally, all ten implants were removed.The nurse still had the central screw within the baseplate and the reversed insert was unattached from the reversed tray.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX42MM STERILE
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER + ADDITIVE, CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16989719
Report Number3000931034-2023-00156
Device Sequence Number6163739
Product Code PAO
UDI-Device Identifier00846832091287
UDI-Public00846832091287
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K201315
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberDWJ742
Device Catalogue NumberDWJ742
Device Lot NumberAE5099008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/27/2023
Supplement Date Received by Manufacturer07/28/2023
Initial Report FDA Received Date05/23/2023
Supplement Report FDA Received Date08/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-