COVIDIEN EVERCROSS 035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number AB35W05200080
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 05/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that an attempt was made to use a evercross to treat a calcified lesion in the proximal superficial femoral artery. degree of tortuosity was little.Degree of calcification was moderate. the vessel was pre-dilated.Embolic protection was used.The balloon was inflated with a syringe.The device was prepped per the ifu, with no issues identified.The device did not pass througha previously-deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used.There were no issues during inflation.No damage noted to the balloon catheter.It is reported that there was balloon deflation difficulties and removal. device would not deflate at the lesion site and there was difficulty removing balloon.Flow was impeded in vessel.The patient had to have surgery to remove part of the balloon that got stuck in the vessel and is due to have further surgery to remove radiopaque marker from the superficial femoral artery.Vessel occlusion due to the device component and vessel spasm were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERCROSS 035
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16990598
Report Number2183870-2023-00186
Device Sequence Number15024249
Product Code LIT
UDI-Device Identifier00643169973787
UDI-Public00643169973787
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K082579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/16/2026
Device Model NumberAB35W05200080
Device Catalogue NumberAB35W05200080
Device Lot NumberB534387
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2023
Initial Report FDA Received Date05/24/2023
Date Device Manufactured03/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-