• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CENTINEL SPINE, LLC PRODISC L SUPERIOR END PLATE MEDIUM 3-STERILE; PROSTHESIS, INTERVERTEBRAL DISC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CENTINEL SPINE, LLC PRODISC L SUPERIOR END PLATE MEDIUM 3-STERILE; PROSTHESIS, INTERVERTEBRAL DISC Back to Search Results
Model Number PDL-M-SP03S
Medical Device Problem Code Expulsion (2933)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/26/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that the patient was implanted with a prodisc l implant on (b)(6) 2023 at l5-s1 level.Approximately two weeks after the implantation surgery the patient reported hearing and feeling a popping noise in her lumbar spine.X-ray images showed that the superior end plate appears to have dislodged anteriorly.The implant was removed on (b)(6) 2023.The pdl device was replaced with a bone alif with screws and washers and subsequent fixation was performed.A review of the dhrs found no anomalies associated with the complaint.Complaint trending found that the rate of complaints is within the levels defined in the risk documentation.A review of the risk assessment found that the risks associated with complaint are identified and mitigated to a level where the benefits outweigh the risks.The removed implant was sent to exponent for device evaluation per their standard pdl evaluation process.There were no anomalies identified during the complaint investigation.The cause for the implant to migrate / dislodge anteriorly is unknown.This report is for 3 of 3 devices involved in this event.
 
Event or Problem Description
The patient was implanted with a prodisc l implant on (b)(6) 2023 at l5-s1 level.Approximately two weeks after the implantation surgery the patient reported hearing and feeling a popping noise in her lumbar spine.X-ray images showed that the superior end plate appears to have dislodged anteriorly.The implant was removed on (b)(6) 2023.The pdl device was replaced with a bone alif with screws and washers and subsequent fixation was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODISC L SUPERIOR END PLATE MEDIUM 3-STERILE
Common Device Name
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
CENTINEL SPINE, LLC
900 airport road, suite 3b
west chester PA 19380
Manufacturer (Section G)
HAMMILL MANUFACTURING COMPANY
360 tomahawk drive
maumee OH 43537
Manufacturer Contact
erin ward
900 airport road
suite 3b
west chester, PA 19380
4848878894
MDR Report Key16991529
Report Number3007494564-2023-00029
Device Sequence Number17016870
Product Code MJO
UDI-Device Identifier00843193114983
UDI-Public00843193114983
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P050010
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPDL-M-SP03S
Device Catalogue NumberN/A
Device Lot Number2022-0810
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2023
Initial Report FDA Received Date05/24/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-