• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EBI, LLC. SPF-XL IIB 2/DM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EBI, LLC. SPF-XL IIB 2/DM Back to Search Results
Model Number N/A
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Date of event: the event occurred sometime in (b)(6) 2023.Medical product: unknown.Therapy date: unknown.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is complete, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient's spf was explanted due to pain.The patient's implant date was (b)(6) 2022 at hmh bayshore medical center.The implanted spf was spf xl iib, 2/dm - 10-1335m, serial number (b)(6).Additional contact and demographic details on the patient are not available.Additional details on the patient's experience are not available.No replacements to our knowledge.The patient stated, "we believe the wire broke after a rear end mva about a year after implantation.My surgeon did take out the battery part of the time.But the cathodes are still in my cervical area.The patient stated that the issue is getting a facility to take an mri needed of my hip and elbow area.The patient was told to contact zimvie.The patient was not sent a replacement product.The product is not expected to be return.No additional information was provided at this time.
 
Event or Problem Description
It was reported that the patient's spf was explanted due to pain.The patient's implant date was on (b)(6) 2022 at (b)(6) medical center.The implanted spf was spf xl iib, 2/dm - 10-1335m, serial number (b)(6).Additional contact and demographic details on the patient are not available.Additional details on the patient's experience are not available.No replacements to our knowledge.The patient stated, "we believe the wire broke after a rear end mva about a year after implantation.My surgeon did take out the battery part of the time.But the cathodes are still in my cervical area.The patient stated that the issue is getting a facility to take an mri needed of my hip and elbow area.The patient was told to contact zimvie.The patient was not sent a replacement product.The product is not expected to be return.No additional patient consequences/ further information has been reported.The spf xl iib was not returned for further evaluation.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The device was not returned to highridge medical for evaluation.The reported event was unable to be confirmed due to limited information received from the customer.The device history record was reviewed, and no discrepancies related to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Highridge medical will continue to monitor for trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPF-XL IIB 2/DM
Common Device Name
SPF
Manufacturer (Section D)
EBI, LLC.
1 gatehall dr
suite 303
parsippany NJ 07054
Manufacturer (Section G)
EBI, LLC
1 gatehall dr
suite 303
parsippany NJ 07054
Manufacturer Contact
drashti patel
1 gatehall dr
suite 303
parsippany, NJ 07054
9732999300
MDR Report Key16991739
Report Number0002242816-2023-00054
Device Sequence Number6815029
Product Code LOE
UDI-Device Identifier00812301020089
UDI-Public00812301020089
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P850035/S023
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberN/A
Device Catalogue Number10-1335M
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2023
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/24/2023
Supplement Report FDA Received Date07/31/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE; SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Disability;
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-