Catalog Number 1-10013-609
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Osseointegration (2646); Implant Pain (4561)
Date of Event 05/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complaint reference number: (b)(4).
 
Event or Problem Description
It was reported that, after left partial knee replacement surgery had been performed on (b)(6) 2022, the patient experienced pain on the left knee.This adverse event was solved by revision surgery performed on (b)(6) 2023, in which all four (4) components were revised, and competitor device were implanted.The engage porous femoral sz 7-lt med had no visible bone ingrowth, the engage porous tibial tray sz 6-lt med had minimal bone ingrowth and the engage tibial insert sz 6-lt med 9mm had visible wear.Current health status of patient is unknown.
 
Additional Manufacturer Narrative
H3, h6: the associated devices were returned and evaluated.The visual inspection revealed scratches, gouges and debris on the devices.The devices show signs of wear.A lab analysis performed on the devices revealed that the anchor stem is missing the tab.Some discoloration of the anchor stem can be seen.The tibial baseplate can be seen.A substance can be seen adhered to the bone contacting porous surface of the tibial baseplate.The femoral component can be seen.A substance can be seen adhered to the bone contacting porous surface of the femoral component.There were no observations of material or manufacturing deviations in the course of this investigation.The clinical/medical investigation concluded that, based on the limited information provided, the reported lack of bony ingrowth was most likely the contributing factor to the knee pain and insert wear; however, a definitive clinical root cause of the insufficient bone ingrowth cannot be concluded.Although field action was generated and a recall decision was made in order to mitigate future recurrence, it is unknown if patient comorbidities may have possibly contributed to the reported event.The patient impact included the reported pain and insert wear likely due to micromotion secondary to lack of bony integration along with the revision/conversion to competitor tka.Further patient impact cannot be determined.No further medical assessment can be rendered at this time.A review of the production records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar previous events, however, no commonalities that would suggest a device deficiency were found.A review of the instructions for use documents for engage partial knee system revealed in the warnings and precautions section that the success of the operation depends on compliance with the operative technique supplied, and the proper use of the instrumentation supplied and specially designed for that range of implants.Additionally, incorrect fixation or positioning of the components, pain and excessive wear of the polyethylene component have been identified in the adverse effects and complications.Deformation or wear from one or more of the prosthetic components is often related to the risk factors detailed in the document.There is not enough data available to conclude that the overall clinical benefit outweighs the potential risk profile when compared to the state of the art.The existing data identifies a potential signal that the performance is an outlier vs the state of the art with respect to the risk for revision.In addition, a historical review concluded that no previous escalated actions for this type of issue were identified.However, as a correction action a voluntary market removal will be performed for the engage cementless partial knee system due recent complaint data that indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, joint tightness or patient condition.Based on this investigation, the need for corrective action is not indicated.
 
Additional Manufacturer Narrative
H10.Additional information in b7 and d10.H11.Corrected information in h6 (health effect - clinical code & health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENGAGE TIBIAL INSERT SZ 6-LT MED 9MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16992262
Report Number1020279-2023-01117
Device Sequence Number6815052
Product Code HSX
UDI-Device Identifier00885556874028
UDI-Public885556874028
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K190439
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number1-10013-609
Device Lot Number100410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2023
Supplement Date Received by Manufacturer10/30/2023
08/12/2024
Initial Report FDA Received Date05/24/2023
Supplement Report FDA Received Date11/07/2023
09/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1673-2023
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1-10011-300 / LOT NUMBER 100274.; 1-10012-600 / LOT NUMBER 100345.; 1-10013-609 / LOT NUMBER 100410.
Patient Age54 YR
Patient SexMale
-
-