Model Number V0100109
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Laceration(s) (1946); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference: (b)(4).
 
Event or Problem Description
It was reported that, during a tka surgery, a femoral distal cutting block from a vis adpt guide lgnp kit cut the knee in valgus.Surgeon had to recut using standard instruments.It is unknown if surgery was delayed due to this event, no patient complications have been reported.
 
Event or Problem Description
It was reported that, during a tka surgery, a femoral distal cutting block from a vis adpt guide lgnp kit cut the knee in valgus.Surgeon had to recut using standard instruments.The procedure was resumed after a non-significant delay, and no further complications have been reported.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation stated that the requested clinical documentation has not been provided for evaluation.It has been communicated via e-mail that no further information is available.Therefore, there were no clinical factors found which would have contributed to the reported event.It was reported the surgeon was ultimately satisfied with the procedure.The patient impact beyond that which already has been reported could not be determined.No further clinical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.As visionaire devices are custom made devices, a review of the complaint history for this part is not applicable.A review of the instructions for use documents for visionaire¿ patient matched instrumentation with fastpak instruments revealed that if the patient matched cutting block does not perform as intended, use the standard smith & nephew instrumentation to complete the surgery.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.A review made by the quality engineering team revealed that after evaluation, no root cause could be found for the complaint.All parts of the design were within visionaire standards.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include surgical technique used or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Additional information: e1 (facility name and address) corrected data: b5, d8 (device available for evaluation?), e1 (initial reporter name), g2, h6 (health effect - clinical code and health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIS ADPT GUIDE LGNP KIT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16992953
Report Number1020279-2023-01119
Device Sequence Number6993595
Product Code JWH
UDI-Device Identifier00885556656396
UDI-Public00885556656396
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K170282
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/26/2023
Device Model NumberV0100109
Device Catalogue NumberV0100109
Device Lot Number00257368V1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2023
Supplement Date Received by Manufacturer07/28/2023
Initial Report FDA Received Date05/24/2023
Supplement Report FDA Received Date08/03/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-