• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR CANNULA; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR CANNULA; CAT Back to Search Results
Model Number OPT316
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
A regulatory body reported, on behalf of a healthcare facility in brazil, that an opt316 optiflow junior nasal cannula had broken.There was no reported patient involvement.
 
Additional Manufacturer Narrative
(b)(4) the optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint opt316 optiflow junior nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information provided by the customer and our knowledge of the product.Result: the customer reported that a opt316 optilow junior nasal cannula was broken.F&p were unable to obtain further information from the customer as it was reported anonmyously to anvisa.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The subject opt316 optiflow junior nasal cannula would have met the required specification at the time of production.The user instructions which accompany the opt316 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.Appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.Do not stretch or crush tube.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR CANNULA
Common Device Name
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key16996798
Report Number9611451-2023-00477
Device Sequence Number20581362
Product Code CAT
UDI-Device Identifier09420012432292
UDI-Public(01)09420012432292(10)2102102089(11)220324
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT316
Device Catalogue NumberOPT316
Device Lot Number2102102089
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2023
Initial Report FDA Received Date05/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-