| Model Number |
500-56150 |
| Medical Device Problem Codes |
Leak/Splash (1354); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that there was a drug leak and occlusion.A 135x50cm ekosonic endovascular device was selected for use in a lower extremity procedure.The catheter was placed, and infusion began.After 9.5 hours of therapy, it was reported that the drug port was leaking at the hub, and blood was coming back up the tubing causing an occlusion.Attempts were made to flush and change the stopcock, but the leak was still occurring.It was decided to discontinue the drip and bring the patient back to reevaluate and attempt to fix the thrombus burden in the vessel.The patient was brought back to the cath lab, and flow was restored using alternative measures that included angiojet and balloon angioplasty.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device eval by manufacturer: the ekosonic endovascular device was returned for analysis.Microscopic visual inspection of the infusion catheter showed cracking on the drug port and coolant luer.Kinking of the infusion catheter was noticed at 16cm and 115 cm from the strain relief.Kinking of the ultrasonic core was noticed at 122cm from the luer barb.The device was tested for leaking, and it was noticed that the device leaked water from the drug port and coolant luer where the cracking was present.The device was unable to be flushed because of the cracked luers causing leakage.
|
| |
|