TORNIER S.A.S. AEQUALIS REVERSED II DIA29 L30 THREADED POST BASE PLATE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWE830
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 04/05/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed. a review of the device history is not possible because the lot number was not communicated. should additional information become available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the rev ii baseplate, screws, and glenosphere were removed because of notching and loosening.
 
Additional Manufacturer Narrative
The reported event could be confirmed.The devices were not returned but evidences were provided, based on provided x-ray, which matches the alleged failure.The provided x-ray (pre-op x-ray) shows the notching of the humeral component on the scapular and the radiolucency around the peripheral screws (present around all screws) and the sphere, indicative of loosening of the baseplate.Since x-ray was provided, the opinion of a medical expert was sought and stated as following: "this case describes notching and loosening of the glenoid component.The x-ray confirms the event.Unfortunately, there is insufficient information (early postoperative x-ray of the index surgery, patient demographics) to make a detailed assessment of a possible root cause".More detailed information about the complaint event as well as the affected devices must be available in order to determine the exact root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that the rev ii baseplate, screws, and glenosphere were removed because of notching and loosening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS REVERSED II DIA29 L30 THREADED POST BASE PLATE
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16997515
Report Number3000931034-2023-00160
Device Sequence Number3377223
Product Code KWS
UDI-Device Identifier03700386952739
UDI-Public03700386952739
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K132285
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberDWE830
Device Catalogue NumberDWE830
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2023
Supplement Date Received by Manufacturer06/27/2023
Initial Report FDA Received Date05/24/2023
Supplement Report FDA Received Date07/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-