MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE PRESERVE STEM 6MM; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE PRESERVE STEM 6MM; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 300-30-06
Medical Device Problem Codes Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Limb Fracture (4518); Multiple Fractures (4519)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via clinical study that the 55 yo female patient experienced a humeral fracture and crack in calcar when broaching for humeral stem.The outcome was last known as continuing.
 
Additional Manufacturer Narrative
D10.Concomitants: 320-36-00 - 36mm humeral liner +0 unconstrained.320-10-00 - equinoxe reverse tray adapter plate tray +0.320-31-36 - glenosphere, 36mm.320-35-02 - small superior augment glenoid plate.
 
Additional Manufacturer Narrative
H3: based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of surgical fracture cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.These devices are used for treatment not diagnosis.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1, h6.Mdr section codes updated/corrected: b, c, d, e, g.Serial number, expiration and manufactured dates unknown.The reason for the intraoperative bone fracture reported cannot be conclusively determined but may be related to a patient condition or high forces during broaching.However, this cannot be confirmed based on the information provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE PRESERVE STEM 6MM
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL
Manufacturer Contact
geoff gannon
MDR Report Key16998670
Report Number1038671-2023-01120
Device Sequence Number12405911
Product Code KWT
UDI-Device Identifier10885862515742
UDI-Public10885862515742
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K162726
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number300-30-06
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2023
Supplement Date Received by Manufacturer04/25/2023
08/28/2024
09/17/2025
Initial Report FDA Received Date05/24/2023
Supplement Report FDA Received Date02/21/2024
09/04/2024
09/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight63 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-