EXACTECH, INC. OPTETRAK; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
CC TIBIAL INSERT SZ 2, 9MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D2b: prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer.D10: concomittants: 204-04-23 - bandeja tib quilla trap cementada 1f/3t, 2f/3t (b)(6); 204-36-08 - vastago ext.16x80mm (b)(6); 204-36-08 - vastago ext.16x80mm (b)(6); 208-01-02 - femur cc nº2 (b)(6); 208-09-07 - adaptador femoral vastago cc 7 grados (b)(6).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
During the week of april 17-25, representatives of exactech, inc.And exactech spain conducted an inperson site visit with (b)(6) hospital.As a result of that site visit and at exactech's request, the hospital provided a retrospective excel listing of revision related to polyethylene wear that have been performed since 2010.These cases have not previously been reported to exactech as complaints.This patient (b)(6)) was implanted with a 208-22-09-inserto tibial cc 2,9mm, serial number (b)(6) on (b)(6) 2014.The insert was manufactured on 7/29/2013 and was packaged in a vacuum bag with no evoh.The insert was manufactured on 7/29/2013 and was 0.62 years shelf age at the time of implant.Pending revision for wear of the polyethylene.No additional details are available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the prosthesis wear reported may have been due to third body wear, patient-related conditions, instability, malalignment between the femoral and tibial components.However, this cannot be confirmed as the devices were not available for evaluation and radiographs were not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial total knee arthroplasty.Subsequently, it was reported that the patient experienced prosthesis wear.As a result, the patient was scheduled for a revision on a future date.No further patient impact reported.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b3, b5, d2b, g1, h6 - health effect - impact code, investigation findings, investigation conclusions, h7 and h11.B3: unknown date.The prosthesis wear reported may have been due to third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components, which led to prosthesis wear.However, this cannot be confirmed as the device was not available for evaluation and images of the explanted device/radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|