MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
| Model Number |
37601 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Communication or Transmission Problem (2896); Insufficient Device Problem Information (3190); Data Problem (3196)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient just got a battery replacement and was programmed with the new dbs clinician tablet.The patient went to the doctor for control and the doctor interrogated the dbs using the old clinician programmer (n'vision).Stimulation was off so the symptom appeared.Once checked with the patient programmer (pp), error 574 appeared.The cause of the error 574 was due to the doctor interrogating with the clinician programmer/n'vision.Using the dbs clinician tablet showed a few error messages/alerts including: low communicator battery, end session: the active group has no valid programs.Stimulation is off., and neurostimulator status: no active group.Stimulation cannot be turned on.The issue was resolved after using the dbs clinician tablet to interrogate the patient, reprogramming session (fulfil the data and put the existing program), and then turn on the stimulation.After stimulation on, the symptom of disease was gone and the patient was receiving effective therapy.They will continue to use the dbs clinician tablet for next programming session.
|
| |
|
Search Alerts/Recalls
|
|
|