Model Number INGENIA 1.5T
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 03/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Philips received a report that a patient suffered a large 2nd degree burn on her thigh after a spinal examination.
 
Additional Manufacturer Narrative
Philips has started an investigation.A follow up will be send when completed h3 other text : incident occurred > 2 months ago, system has been in use since without issues.
 
Additional Manufacturer Narrative
Conclusion: based on the available information, there is no indication of a malfunction of the mr system or used inbuilt posterior coil that could have contributed to the incident.The observed reddening and blistering on the images provided can be explained by direct contact between the patient and the bore wall.Even though it was mentioned that soft padding was used to prevent bore contact, considering the patients size and duration of the examination (half hour) it is highly likely that either there was insufficient padding or that the padding was moved.The reported burn is considered a second degree burn and was determined to be a serious injury.Contributing factors: the patient was obese, suffered from diabetes and had a fever.The thermoregulation of obese and diabetic patients is known to be impaired.The risk of rf energy-related injuries is higher in patients with impaired thermoregulation.The total administered specific energy dose of 2.3 kj/kg is high for a patient with impaired thermoregulation.The instructions for use advises to preferably keep sed <2.0 kj/kg for those patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIA 1.5T
Common Device Name
SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
PHILIPS ELECTRONICS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS ELECTRONICS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs street
cambridge, MA 02141
MDR Report Key17001754
Report Number3003768277-2023-03299
Device Sequence Number6579938
Product Code LNH
UDI-Device Identifier00884838009820
UDI-Public884838009820
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K110151
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberINGENIA 1.5T
Device Catalogue Number781396
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2023
Supplement Date Received by Manufacturer05/17/2023
Initial Report FDA Received Date05/25/2023
Supplement Report FDA Received Date07/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age70 YR
Patient SexFemale
Patient Weight160 KG
-
-