Catalog Number 999890249 |
Device Problems
Degraded (1153); Naturally Worn (2988)
|
Patient Problems
Pain (1994); Synovitis (2094); Metal Related Pathology (4530)
|
Event Date 11/07/2022 |
Event Type
Injury
|
Event Description
|
On (b)(6) 2007, the patient undergone right total hip arthroplasty with a depuy products.On (b)(6) 2022, the patient had a revision of failed right total hip arthroplasty, acetabular component and head.Elevated cobalt chromium levels.Operative noted corrosion on the femoral head with fair amount of black debris from the metal-on-metal and synovitis was also removed.Acetabular component also noted with bone loss.Doi: (b)(6) 2007, dor: (b)(6) 2022, affected side: right hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).E3 initial reporter occupation: lawyer.H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Plaintiff fact sheet received.Litigation record alleges that patient experienced hip pain when standing or sitting and chromium and cobalt were testes high.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|