| Model Number |
SIGTRS60AXT |
| Medical Device Problem Codes |
Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic pancreatic tail resection, during resecting the pancreas after compressions, while using a powered stapler, the pancreas was clamped, the green button did not light up, and the firing could not be performed.The pancreas had been pre-compressed and was quite thin, and there was no indication that it was out of indicated applicable range.After that, all components, including the cartridge were removed, but the lcd screen of the handle remained in firing preparation complete state, and the handle was in a frozen state.The surgeon used a manual stapler with a new cartridge to resolve the issue.After that, the handle was then restarted with a power reset, and the device started up normally.The surgical procedure was extended by more than 30 minutes with no adverse event to the patient.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product/s: sigphandle sig power sigphandle handle, serial #:(b)(6); sigpshell sig power sigpshell control shell, lot #:unknown; sigadaptstnd sig power sigadaptstnd linear adapter, serial #:unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|