• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS Back to Search Results
Model Number 550AVHCT
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Per the manufacturer's subsidiary's clinical specialist, the deviation was not found in any of the measured parameters, only the calculated be value.He found that the calculation formula of the bpm was different than that of the bga.
 
Event or Problem Description
It was reported that during use of the device for cardiopulmonary bypass (cpb), the base excess (be) values on the blood parameter monitor (bpm) were deviating from the values gathered by the blood gas analyzer (bga).As mitigation, the values from the bga were used.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Event or Problem Description
Per clinical review: the user reported an issue with their blood parameter monitor (bpm) during cardiopulmonary bypass (cpb).Specifically, the user observed a deviation of the base excess value when compared to the values obtained from the lab.It is unclear if this deviation is a positive or negative deviation from the lab blood gas values.There was no delay to the procedure.The unit was not changed out, and the surgery was completed successfully.There was no blood loss or adverse events reported.The user confirmed base excess (be) values from an abl 90 flex blood gas analyzer.
 
Additional Manufacturer Narrative
The reported complaint was not verifiable since the product was not returned for evaluation or testing.Multiple diligence attempts for part return were unsuccessful.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 550
Common Device Name
CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson raod
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key17003033
Report Number1828100-2023-00150
Device Sequence Number15245926
Product Code DRY
UDI-Device Identifier00886799001783
UDI-Public(01)00886799001783(11)201024
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K182110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number550AVHCT
Device Catalogue Number550AVHCT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2023
Supplement Date Received by Manufacturer05/30/2023
07/21/2023
Initial Report FDA Received Date05/25/2023
Supplement Report FDA Received Date06/20/2023
08/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/24/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABL 90 FLEX BLOOD GAS ANALYZER.
Patient SexUnknown
-
-