| Model Number |
IPN917235 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during use on a patient with injection pressure at 500psi, the catheter ruptured.As a result, the catheter was removed a 2nd catheter was inserted at the same insertion site.No patient harm or injury.When the ruptured occurred, the cardiologist was struck in the eye with contrast.Additional information received states that there was no harm or injury to the cardiologist.The contrast agent was not mixed with any patient blood or bodily fluid, protective eyewear was being worn at the time of the event, and the cardiologist did not require to be seen by occupational health or need further medical evaluation or treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Other remarks: n/a.Corrected data: n/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#: (b)(4).The reported lot number (16f22l0045) matches the lot number on the returned original packaging pouch/label.Returned for investigation was a 5fr.80cm berman catheter with the original packaging pouch.The sample was returned in a cardboard box and was in a sealed bio-hazard bag.Upon return, the catheter body was immediately noted ruptured near the junction hub; the body was noted ruptured from approximately 88.2cm to 88.3cm from the distal tip of the catheter.The supplied control stroke syringe was noted connected to the inflation lumen stopcock; no damage or abnormalities were noted to the returned syringe.The inflation lumen stopcock was in the open position.The recommended volume capacity of the balloon is 0.75cc.Upon microscopic inspection, the balloon appeared typical; no damage or abnormalities were noted to the balloon.No condensation was noted within the inflation lumen extension line.Some dried contrast media was noted within the injection lumen extension line.Spots of dried blood/contrast media was noted on the exterior surfaces of the returned sample.No other damage or abnormalities were noted.The inflation lumen was injected with 0.75cc of air using the returned control stroke syringe.The balloon inflated symmetrically.One side of the balloon measured approximately 4mm.The other side measured approximately 4mm.The balloon did meet specifications per graphic i-07135-003e rev.00 of radius ratio less than or equal to 2.0.The inflation lumen was injected with 0.75cc of air using the returned control stroke syringe.The balloon inflated symmetrically.The balloon deflated in less than 3 seconds when the syringe was removed per specification i-07135-003e rev.00.No pull away was noted after the tug test.The balloon was placed in water and air was injected into the inflation lumen again.No leak was noted.The ruptured catheter body did not result in damage to the inflation lumen.The catheter's injection lumen was aspirated/flushed , and air was noted leaking from the ruptured catheter body.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint of catheter body ruptured in use is confirmed.The catheter body was found ruptured near the junction hub during visual inspection of the returned sample.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the ruptured catheter body.The probable root cause of the complaint is manufacturing related.A corrective and preventive action has been initiated to address the issue.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that during use on a patient with injection pressure at 500psi, the catheter ruptured.As a result, the catheter was removed a 2nd catheter was inserted at the same insertion site.No patient harm or injury.When the ruptured occurred, the cardiologist was struck in the eye with contrast.Additional information received states that there was no harm or injury to the cardiologist.The contrast agent was not mixed with any patient blood or bodily fluid, protective eyewear was being worn at the time of the event, and the cardiologist did not require to be seen by occupational health or need further medical evaluation or treatment.
|
| |
|
Search Alerts/Recalls
|