| Model Number |
SO-1968 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Decreased Respiratory Rate (2485)
|
| Date of Event |
01/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reportable as the patient desaturated.
|
| |
|
Event or Problem Description
|
|
Very clear cuts located 4 different places on the o2 tubing w/connector the patient desaturated while plugged into his tank while the saturation was good when plugged into the laptop.The tubing was discovered with very clear cuts of 1 to 2 cm located in 4 different places over the last 2 meters at one end.
|
| |
|
Additional Manufacturer Narrative
|
|
Reportable as the patient desaturated.Complaint history reviewed.No similar complaints for product under so-1968 or 2030 in the previous 24 months.Without further information, complaint cannot be investigated further.Most likely root cause is product was cut by a sharp object or corner during use.Risk(rma-20017c): r88: low oxygen delivery - tubing breaks or unravels - s=6, o=3, rpn=18.Rpn < 25, therefore risk is acceptable.
|
| |
|
Event or Problem Description
|
|
Very clear cuts located 4 different places on the o2 tubing w/connector the patient desaturated while plugged into his tank while the saturation was good when plugged into the laptop.The tubing was discovered with very clear cuts of 1 to 2 cm located in 4 different places over the last 2 meters at one end.
|
| |
|