| Model Number |
0502503010 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not received for investigation at stryker endoscopy because it was repaired locally in stryker australia.The technical service report indicates: fault diagnosis and results.Diagnosis: on examination of the scope, the internal glass rod has been shattered and is causing disruption to image quality.A replacement will be supplied.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|